台灣開源呼吸機研究團隊

緣起

看到了這個專案,順便分享

社團名稱

中文: 台灣開源呼吸機研究團隊
English: OSMS-Taiwan Ventilator Team (OSMSTW-MIT)

注意事項

醫療用途的設備不能隨意買賣,此處純粹研究製作方法。
任何測試與驗證,請勿直接使用在人身上

文件

版權以 CC0 釋出
目前的文件狀態都是草稿,夥伴可以隨意補充,待討論到了比較成熟的時候,我們會在會議上部分確認,形成共識

大事記

相關文件


原文

Updated 28 March 2020
Note changes to specifications with increased BPM and I:E ratios. Adjusting one of these parameters may influence the other, since at higher BPM, the tidal volume is usually decreased. Consult a clinician.
From the Clinical Guidance this document summarizes the minimum set of requirements for ventilation:
  1. Patients must be under the management of a trained clinician.
  1. The minimum controllable parameters in order to ventilate a patient include:
  1. Respiratory Rate (RR) (breaths per minute): between 8 – 40
  1. Tidal Volume (TV) (air volume pushed into lung): between 200 – 800 mL based on patient weight
  1. I/E Ratio (inspiratory/expiration time ratio): recommended to start around 1:2; best if adjustable between range of 1:1 – 1:4*
  1. Assist Detection pressure. When a patient tries to inspire, they can cause a dip on the order of 1 – 5 cm H2O, with respect to PEEP pressure (not necessarily = atmospheric).
  1. Airway pressure must be monitored
  1. Maximum pressure should be limited to 40 cm H2O at any time; Plateau pressure should be limited to max 30 cm H2O
  1. The use of a passive mechanical blow-off valve fixed at 40 cm H2O is strongly recommended
  1. Clinician require readings of plateau pressure and PEEP (refer to clinical documentation tab)
  1. PEEP of 5-15 cm H2O required; many patients need 10-15 cmH2O
  1. Failure conditions must permit conversion to manual clinician override, i.e. if automatic ventilation fails, the conversion to immediate ventilation must be immediate.
  1. Ventilation on room air is better than no ventilation at all. Blending of oxygen and air gas mixture to adjust FiO2 is not important in an emergency scenario.  It is certainly nice to have that ability and can easily be implemented with an oxygen / air gas blender that some hospitals already have.
  1. COVID-19 can get aerosolized (airborne), so HEPA filtration on the patient’s exhalation is required or between the ventilator unit and the patient (at the end of the endotracheal tube) to protect clinical staff from certain infection. In-line HEPA filters can usually b
上次編輯:wuulong sheu (5 個月前)

基本資料

  • 目前原網站有規格與設計公開,但似乎原始碼還沒開放下載。BOM 似乎也不夠清楚
MIT 下載區

專案